Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations
computed tomography
multi-detector CT (MDCT)
medical radiation dose
image quality
Bibliogr. s. e66-e67
Purpose: To assess and compare radiation dose and image quality from non-contrast head and neck computed tomography (CT) examinations from four different multi-detector CT (MDCT) scanners. Material and methods: Four CT scanners with different numbers of detector rows including one 4-MDCT, a 6-MDCT, a 16-MDCT, and a 64-MDCT were investigated. Common CT dose descriptors including volumetric CT dose index (CTDIv), dose length product (DLP), and the effective dose (ED), and image quality parameters include image noise, uniformity, and spatial resolution (SR) were estimated for each CT scanner with standard tools and methods. To have a precise comparison between CT scanners and related doses and image quality parameters, the ImPACT Q-factor was used. Results: Minimum and maximum CTDIv, DLP, and ED in the head scan were 18 ± 3 and 49 ± 4 mGy, 242 ± 28 and 692 ± 173 mGy × cm, 0.46 ± 0.4 and 1.31 ± 0.33 mSv for 16-MDCT and 64-MDCT, respectively. And 16 ± 2 to 27 ± 3, 286 ± 127 to 645 ± 79 and 1.46 ± 0.65 to 3.29 ± 0.40 for neck scan, respectively. The Q-factor in head scan was 2.4, 3.3, 4.4 and 5.6 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. The Q-factor in neck scan was 3.4, 4.6, 4.7 and 6.0 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. Conclusions: The results clearly indicate an increasing trend in the Q-factor from 4-MDCT to 64-MDCT units in both head and neck examinations. This increasing trend is due to a better SR and less noise of images taken and/or fewer doses in 64-MDCT.
| cris.lastimport.wos | 2024-04-09T20:40:43Z | |
| dc.abstract.en | Purpose: To assess and compare radiation dose and image quality from non-contrast head and neck computed tomography (CT) examinations from four different multi-detector CT (MDCT) scanners. Material and methods: Four CT scanners with different numbers of detector rows including one 4-MDCT, a 6-MDCT, a 16-MDCT, and a 64-MDCT were investigated. Common CT dose descriptors including volumetric CT dose index (CTDIv), dose length product (DLP), and the effective dose (ED), and image quality parameters include image noise, uniformity, and spatial resolution (SR) were estimated for each CT scanner with standard tools and methods. To have a precise comparison between CT scanners and related doses and image quality parameters, the ImPACT Q-factor was used. Results: Minimum and maximum CTDIv, DLP, and ED in the head scan were 18 ± 3 and 49 ± 4 mGy, 242 ± 28 and 692 ± 173 mGy × cm, 0.46 ± 0.4 and 1.31 ± 0.33 mSv for 16-MDCT and 64-MDCT, respectively. And 16 ± 2 to 27 ± 3, 286 ± 127 to 645 ± 79 and 1.46 ± 0.65 to 3.29 ± 0.40 for neck scan, respectively. The Q-factor in head scan was 2.4, 3.3, 4.4 and 5.6 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. The Q-factor in neck scan was 3.4, 4.6, 4.7 and 6.0 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. Conclusions: The results clearly indicate an increasing trend in the Q-factor from 4-MDCT to 64-MDCT units in both head and neck examinations. This increasing trend is due to a better SR and less noise of images taken and/or fewer doses in 64-MDCT. | pl |
| dc.contributor.author | Khoramian, Daryoush | pl |
| dc.contributor.author | Sistani, Soroush | pl |
| dc.contributor.author | Firouzjah, Razzagh Abedi | pl |
| dc.date.accessioned | 2019-02-19T11:37:37Z | |
| dc.date.available | 2019-02-19T11:37:37Z | |
| dc.date.issued | 2019 | pl |
| dc.date.openaccess | 0 | |
| dc.description.accesstime | w momencie opublikowania | |
| dc.description.additional | Bibliogr. s. e66-e67 | pl |
| dc.description.physical | e61-e67 | pl |
| dc.description.version | ostateczna wersja wydawcy | |
| dc.description.volume | 84 | pl |
| dc.identifier.doi | 10.5114/pjr.2019.82743 | pl |
| dc.identifier.eissn | 1899-0967 | pl |
| dc.identifier.issn | 1733-134X | pl |
| dc.identifier.project | ROD UJ / OP | pl |
| dc.identifier.uri | https://ruj.uj.edu.pl/xmlui/handle/item/68731 | |
| dc.language | eng | pl |
| dc.language.container | eng | pl |
| dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 4.0 Międzynarodowa | * |
| dc.rights.licence | CC-BY-NC-ND | |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/legalcode.pl | * |
| dc.share.type | otwarte czasopismo | |
| dc.source.integrator | false | |
| dc.subject.en | computed tomography | pl |
| dc.subject.en | multi-detector CT (MDCT) | pl |
| dc.subject.en | medical radiation dose | pl |
| dc.subject.en | image quality | pl |
| dc.subtype | Article | pl |
| dc.title | Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations | pl |
| dc.title.journal | Polish Journal of Radiology | pl |
| dc.type | JournalArticle | pl |
| dspace.entity.type | Publication |
* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.
Views
9
Views per month
Views per city
Downloads
Open Access