Purpose: There are two standard methods for an magnetic resonance imaging (MRI) examination of the sternoclavicular joints: with loop coils and the patient in the prone position, or with torso coils with the patient in a supine position. In some centres these joints are examined with the spine coil in a patient laying prone. There are no reports on the advantages of this method. Our hypothesis is that despite different MRI systems, application of a spine coil will improve examination quality. Material and methods: Twenty-one healthy volunteers (10 female, 11 male, mean age 25 years) were randomised into three groups and scanned using three different MRI scanners (1.5T: Siemens Avanto, Philips Ingenia, 3.0T: Philips Achieva). Each volunteer was examined twice: using a standard protocol with a torso coil and with a spine coil, in prone position. The two groups were compared with regard to the intensity of motion artefacts using the χ2 test, and to the signal-to-noise ratio with the Wilcoxon signed-rank test. Results: Application of a spine coil resulted in a significant decrease in the number of motion artefacts in all three planes (axial: p = 0.0004; sagittal: p < 0.0001; coronal: p = 0.0054). Moreover, the signal-to-noise ratio was significantly increased with the application of a spine coil (28.6 ± 8.6 vs. 18.5 ± 7.3, respectively; p = 0.0002). Conclusions: Application of a spine coil with the patient in a prone position is suitable for MRI evaluation of the sternoclavicular joints. It allows a higher signal-to-noise ratio and a lower intensity of motion artefacts to be obtained compared to a torso coil.
cris.lastimport.wos | 2024-04-10T03:06:55Z | |
dc.abstract.en | Purpose: There are two standard methods for an magnetic resonance imaging (MRI) examination of the sternoclavicular joints: with loop coils and the patient in the prone position, or with torso coils with the patient in a supine position. In some centres these joints are examined with the spine coil in a patient laying prone. There are no reports on the advantages of this method. Our hypothesis is that despite different MRI systems, application of a spine coil will improve examination quality. Material and methods: Twenty-one healthy volunteers (10 female, 11 male, mean age 25 years) were randomised into three groups and scanned using three different MRI scanners (1.5T: Siemens Avanto, Philips Ingenia, 3.0T: Philips Achieva). Each volunteer was examined twice: using a standard protocol with a torso coil and with a spine coil, in prone position. The two groups were compared with regard to the intensity of motion artefacts using the χ2 test, and to the signal-to-noise ratio with the Wilcoxon signed-rank test. Results: Application of a spine coil resulted in a significant decrease in the number of motion artefacts in all three planes (axial: p = 0.0004; sagittal: p < 0.0001; coronal: p = 0.0054). Moreover, the signal-to-noise ratio was significantly increased with the application of a spine coil (28.6 ± 8.6 vs. 18.5 ± 7.3, respectively; p = 0.0002). Conclusions: Application of a spine coil with the patient in a prone position is suitable for MRI evaluation of the sternoclavicular joints. It allows a higher signal-to-noise ratio and a lower intensity of motion artefacts to be obtained compared to a torso coil. | pl |
dc.contributor.author | Kusak, Artur Stefan | pl |
dc.contributor.author | Podgórski, Michał Tomasz | pl |
dc.contributor.author | Grzelak, Piotr | pl |
dc.contributor.author | Kwapisz, Adam | pl |
dc.date.accessioned | 2019-02-15T14:02:31Z | |
dc.date.available | 2019-02-15T14:02:31Z | |
dc.date.issued | 2018 | pl |
dc.date.openaccess | 0 | |
dc.description.accesstime | w momencie opublikowania | |
dc.description.additional | Bibliogr. s. e518 | pl |
dc.description.physical | e514-e518 | pl |
dc.description.version | ostateczna wersja wydawcy | |
dc.description.volume | 83 | pl |
dc.identifier.doi | 10.5114/pjr.2018.81674 | pl |
dc.identifier.eissn | 1899-0967 | pl |
dc.identifier.issn | 1733-134X | pl |
dc.identifier.project | ROD UJ / OP | pl |
dc.identifier.uri | https://ruj.uj.edu.pl/xmlui/handle/item/68508 | |
dc.language | eng | pl |
dc.language.container | eng | pl |
dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 4.0 Międzynarodowa | * |
dc.rights.licence | CC-BY-NC-ND | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/legalcode.pl | * |
dc.share.type | otwarte czasopismo | |
dc.subject.en | MRI | pl |
dc.subject.en | technique | pl |
dc.subject.en | protocol | pl |
dc.subject.en | sternoclavicular joints | pl |
dc.subject.en | torso coil | pl |
dc.subject.en | spine coil | pl |
dc.subtype | Article | pl |
dc.title | The advantages of a spine coil over a torso coil in magnetic resonance imaging examination of the sternoclavicular joints | pl |
dc.title.journal | Polish Journal of Radiology | pl |