The assessment of diagnostic medical images using 10-bit grayscale : fact or myth?

2018
journal article
article
cris.lastimport.wos2024-04-09T21:34:47Z
dc.abstract.enPurpose: The aim of the study was to find tests for verification of 10-bit grayscale support for radiological workstation hardware/operating system and for DICOM viewer. Also, we wanted to perform the tests on available workstations. Material and methods: The following tests were selected to verify 10-bit grayscale support: workstation hardware and operating system tests using DirectX SDK 10BitScanout10.exe and NEC 10bitdemo.exe programs, DICOM viewer tests using TG18-MP and TG18-QC patterns. The tests were performed on 14 radiological workstations in a local department of radiology and on 4 radiological workstations during the technical exhibition at the ECR (European Congress of Radiology) in 2017. Results: Only 3 out of 14 local workstations passed the hardware and operating system 10-bit grayscale support test. Only one workstation passed the DICOM viewer 10-bit grayscale support test. Conclusions: Hardware/operating system and DICOM viewer 10-bit grayscale support should be tested on every radiological workstation even if compatibility is stated by the manufacturer. Existing medical diagnostic display quality requirements should be adjusted so that 10-bit support for DICOM viewers is only a desirable option.pl
dc.affiliationWydział Lekarski : Zakład Radiologiipl
dc.cm.id89793
dc.contributor.authorChrzan, Robert - 129018 pl
dc.contributor.authorUrbanik, Andrzej - 133713 pl
dc.date.accessioned2018-05-28T15:43:58Z
dc.date.available2018-05-28T15:43:58Z
dc.date.issued2018pl
dc.date.openaccess0
dc.description.accesstimew momencie opublikowania
dc.description.additionalBibliogr. s. e131-e132pl
dc.description.physicale127-e132pl
dc.description.points15pl
dc.description.versionostateczna wersja wydawcy
dc.description.volume83pl
dc.identifier.doi10.5114/pjr.2018.75877pl
dc.identifier.eissn1899-0967pl
dc.identifier.issn1733-134Xpl
dc.identifier.projectROD UJ / Ppl
dc.identifier.urihttps://ruj.uj.edu.pl/xmlui/handle/item/54609
dc.languageengpl
dc.language.containerengpl
dc.rightsUdzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska*
dc.rights.licenceCC-BY-NC-ND
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode*
dc.share.typeotwarte czasopismo
dc.subject.endata displaypl
dc.subject.endiagnostic imagingpl
dc.subject.enquality controlpl
dc.subtypeArticlepl
dc.titleThe assessment of diagnostic medical images using 10-bit grayscale : fact or myth?pl
dc.title.journalPolish Journal of Radiologypl
dc.typeJournalArticlepl
dspace.entity.typePublication
cris.lastimport.wos
2024-04-09T21:34:47Z
dc.abstract.enpl
Purpose: The aim of the study was to find tests for verification of 10-bit grayscale support for radiological workstation hardware/operating system and for DICOM viewer. Also, we wanted to perform the tests on available workstations. Material and methods: The following tests were selected to verify 10-bit grayscale support: workstation hardware and operating system tests using DirectX SDK 10BitScanout10.exe and NEC 10bitdemo.exe programs, DICOM viewer tests using TG18-MP and TG18-QC patterns. The tests were performed on 14 radiological workstations in a local department of radiology and on 4 radiological workstations during the technical exhibition at the ECR (European Congress of Radiology) in 2017. Results: Only 3 out of 14 local workstations passed the hardware and operating system 10-bit grayscale support test. Only one workstation passed the DICOM viewer 10-bit grayscale support test. Conclusions: Hardware/operating system and DICOM viewer 10-bit grayscale support should be tested on every radiological workstation even if compatibility is stated by the manufacturer. Existing medical diagnostic display quality requirements should be adjusted so that 10-bit support for DICOM viewers is only a desirable option.
dc.affiliationpl
Wydział Lekarski : Zakład Radiologii
dc.cm.id
89793
dc.contributor.authorpl
Chrzan, Robert - 129018
dc.contributor.authorpl
Urbanik, Andrzej - 133713
dc.date.accessioned
2018-05-28T15:43:58Z
dc.date.available
2018-05-28T15:43:58Z
dc.date.issuedpl
2018
dc.date.openaccess
0
dc.description.accesstime
w momencie opublikowania
dc.description.additionalpl
Bibliogr. s. e131-e132
dc.description.physicalpl
e127-e132
dc.description.pointspl
15
dc.description.version
ostateczna wersja wydawcy
dc.description.volumepl
83
dc.identifier.doipl
10.5114/pjr.2018.75877
dc.identifier.eissnpl
1899-0967
dc.identifier.issnpl
1733-134X
dc.identifier.projectpl
ROD UJ / P
dc.identifier.uri
https://ruj.uj.edu.pl/xmlui/handle/item/54609
dc.languagepl
eng
dc.language.containerpl
eng
dc.rights*
Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska
dc.rights.licence
CC-BY-NC-ND
dc.rights.uri*
http://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode
dc.share.type
otwarte czasopismo
dc.subject.enpl
data display
dc.subject.enpl
diagnostic imaging
dc.subject.enpl
quality control
dc.subtypepl
Article
dc.titlepl
The assessment of diagnostic medical images using 10-bit grayscale : fact or myth?
dc.title.journalpl
Polish Journal of Radiology
dc.typepl
JournalArticle
dspace.entity.type
Publication
Affiliations

* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.

Views
8
Views per month
Views per city
Hong Kong
7
Ashburn
1
Downloads
chrzan_urbanik_the_assessment_of_diagnostic_medical_images_using_10-bit_grayscale_2018.pdf
26