Dose uncertainty due to energy dependence in dual-energy computed tomography
dual-energy CT
energy absorption coefficient
Bibliogr. s. e274
Purpose: To evaluate the absolute dose uncertainty at 2 different energies and for the large and small bowtie filters in dual-energy computed tomography (DECT). Material and methods: Measurements were performed using DECT at 80 kV and 140 kilovoltage peak (kVp), and single-energy computed tomography (CT) at 120 kV. The absolute dose was calculated from the mass-energy absorption obtained from the half-value layer (HVL) of aluminium. Results: The difference in the water-to-air ratio of the mean mass energy-absorption coefficients at 80 kV and 140 kV was 2.0% for the small bow-tie filter and 3.0% for the large bow-tie filter. At lower tube voltages, the difference in the absorbed dose with the large and small bow-tie filters was larger. Conclusions: The absolute dose uncertainty due to energy dependence was 3.0%, which could be reduced with single-energy beams at 120 kV or by using the average effective energy measurement with dual-energy beams.
cris.lastimport.wos | 2024-04-09T21:10:31Z | |
dc.abstract.en | Purpose: To evaluate the absolute dose uncertainty at 2 different energies and for the large and small bowtie filters in dual-energy computed tomography (DECT). Material and methods: Measurements were performed using DECT at 80 kV and 140 kilovoltage peak (kVp), and single-energy computed tomography (CT) at 120 kV. The absolute dose was calculated from the mass-energy absorption obtained from the half-value layer (HVL) of aluminium. Results: The difference in the water-to-air ratio of the mean mass energy-absorption coefficients at 80 kV and 140 kV was 2.0% for the small bow-tie filter and 3.0% for the large bow-tie filter. At lower tube voltages, the difference in the absorbed dose with the large and small bow-tie filters was larger. Conclusions: The absolute dose uncertainty due to energy dependence was 3.0%, which could be reduced with single-energy beams at 120 kV or by using the average effective energy measurement with dual-energy beams. | pl |
dc.contributor.author | Kawahara, Daisuke | pl |
dc.contributor.author | Toyoda, Takatoshi | pl |
dc.contributor.author | Yokomachi, Kazushi | pl |
dc.contributor.author | Fujioka, Chikako | pl |
dc.contributor.author | Nagata, Yasushi | pl |
dc.date.accessioned | 2023-07-26T08:44:05Z | |
dc.date.available | 2023-07-26T08:44:05Z | |
dc.date.issued | 2023 | pl |
dc.date.openaccess | 0 | |
dc.description.accesstime | w momencie opublikowania | |
dc.description.additional | Bibliogr. s. e274 | pl |
dc.description.physical | e270-e274 | pl |
dc.description.version | ostateczna wersja wydawcy | |
dc.description.volume | 88 | pl |
dc.identifier.doi | 10.5114/pjr.2023.128682 | pl |
dc.identifier.eissn | 1899-0967 | pl |
dc.identifier.issn | 1733-134X | pl |
dc.identifier.uri | https://ruj.uj.edu.pl/xmlui/handle/item/317169 | |
dc.language | eng | pl |
dc.language.container | eng | pl |
dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 4.0 Międzynarodowa | * |
dc.rights.licence | CC-BY-NC-ND | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/legalcode.pl | * |
dc.share.type | otwarte czasopismo | |
dc.subject.en | dual-energy CT | pl |
dc.subject.en | energy absorption coefficient | pl |
dc.subtype | Article | pl |
dc.title | Dose uncertainty due to energy dependence in dual-energy computed tomography | pl |
dc.title.journal | Polish Journal of Radiology | pl |
dc.type | JournalArticle | pl |
dspace.entity.type | Publication |
* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.
Views
7
Views per month
Views per city
Downloads
Open Access