Simple view
Full metadata view
Authors
Statistics
Resistor-less charge sensitive amplifier for semiconductor detectors
front-end electronics for detector readout
gamma spectroscopy
resistor-less charge sensitive amplifier
ultra-low background
A new concept of a Charge Sensitive Amplifier without a high-value resistor in the feedback loop is presented. Basic spectroscopic parameters of the amplifier coupled to a coaxial High Purity Germanium detector (HPGe) are discussed. The amplifier signal input is realized with an n-channel J-FET transistor. The feedback capacitor is discharged continuously by the second, forward biased n-channel J-FET, driven by an RC low–pass filter. Both the analog—with a standard spectroscopy amplifier and a multi-channel analyzer—and the digital—by applying a Flash Analog to Digital Converter—signal readouts were tested. The achieved resolution in the analog and the digital readouts was 0.17% and 0.21%, respectively, at the Full Width at Half Maximum of the registered 60Co 1332.5 keV gamma line.
| dc.abstract.en | A new concept of a Charge Sensitive Amplifier without a high-value resistor in the feedback loop is presented. Basic spectroscopic parameters of the amplifier coupled to a coaxial High Purity Germanium detector (HPGe) are discussed. The amplifier signal input is realized with an n-channel J-FET transistor. The feedback capacitor is discharged continuously by the second, forward biased n-channel J-FET, driven by an RC low–pass filter. Both the analog—with a standard spectroscopy amplifier and a multi-channel analyzer—and the digital—by applying a Flash Analog to Digital Converter—signal readouts were tested. The achieved resolution in the analog and the digital readouts was 0.17% and 0.21%, respectively, at the Full Width at Half Maximum of the registered 60Co 1332.5 keV gamma line. | pl |
| dc.affiliation | Wydział Fizyki, Astronomii i Informatyki Stosowanej : Instytut Fizyki im. Mariana Smoluchowskiego | pl |
| dc.contributor.author | Pelczar, Krzysztof - 103639 | pl |
| dc.contributor.author | Panas, Krzysztof - 241600 | pl |
| dc.contributor.author | Zuzel, Grzegorz - 186344 | pl |
| dc.date.accessioned | 2016-11-22T10:48:27Z | |
| dc.date.available | 2016-11-22T10:48:27Z | |
| dc.date.issued | 2016 | pl |
| dc.description.physical | 142-147 | pl |
| dc.description.volume | 835 | pl |
| dc.identifier.doi | 10.1016/j.nima.2016.08.036 | pl |
| dc.identifier.eissn | 1872-9576 | pl |
| dc.identifier.issn | 0168-9002 | pl |
| dc.identifier.uri | http://ruj.uj.edu.pl/xmlui/handle/item/32611 | |
| dc.language | eng | pl |
| dc.language.container | eng | pl |
| dc.rights | Dodaję tylko opis bibliograficzny | * |
| dc.rights.licence | Bez licencji otwartego dostępu | |
| dc.source.integrator | false | |
| dc.subject.en | front-end electronics for detector readout | pl |
| dc.subject.en | gamma spectroscopy | pl |
| dc.subject.en | resistor-less charge sensitive amplifier | pl |
| dc.subject.en | ultra-low background | pl |
| dc.subtype | Article | pl |
| dc.title | Resistor-less charge sensitive amplifier for semiconductor detectors | pl |
| dc.title.journal | Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment | pl |
| dc.type | JournalArticle | pl |
| dspace.entity.type | Publication |