Individual doses in screening mammography examinations performed with digital equipment in Poland

2011
journal article
other documents
dc.abstract.enBackground: The aim of this study was the estimation of individual doses for women participating in the mammography screening program in Poland, examined with digital systems. Material/Methods: Materials for this study included data obtained from 1800 exposures performed with the use of 9 different mammography units with digital systems. Exposure conditions (breast thickness after compression, high voltage, mAs value, the material of the anode and additional filter) were recorded during mammography screening examinations conducted for over one year in Poland. The collected data were used to determine individual doses in accordance with the method, which uses conversion factors for air kerma to the absorbed dose in the glandular tissue. The entrance air kerma for compressed breast exposure was determined on the basis of the measurements of the air kerma for the high-voltage value at which the exposure was performed, assuming a linear air kerma dependence on the tube load. Results: Individual doses established in 9 mammography facilities ranged from 0.40 mGy to 11.48 mGy for CR (computed radiography) systems and from 0.36 mGy to 5.80 mGy for DR (digital radiography) systems. The maximum value of the individual doses for CR systems was about 98% higher than for DR systems. The values of compressed breast thickness ranged from 2.0 cm to 7.6 cm for CR systems and from 2.1 cm to 11.4 cm for DR systems. The glandularity of breasts was from 9% to 100% for CR systems and from 3% to 97% for DR systems. The mean glandularity of breasts was 47% for CR systems and 29% for DR systems. Conclusions: One of the criteria for evaluating the suitability of digital systems in breast cancer screening in Poland should be the systematic evaluation of individual doses. In Poland, it is necessary to implement standardized methods of optimization of the exposure parameters for reducing individual doses to a minimum necessary to maintain the image quality.pl
dc.contributor.authorFabiszewska, Ewapl
dc.contributor.authorGrabska, Iwonapl
dc.contributor.authorPasicz, Katarzynapl
dc.date.accession2018-09-18pl
dc.date.accessioned2018-09-18T10:57:04Z
dc.date.available2018-09-18T10:57:04Z
dc.date.issued2011pl
dc.date.openaccess0
dc.description.accesstimew momencie opublikowania
dc.description.additionalBibliogr. s. 79-80pl
dc.description.number4pl
dc.description.physical75-80pl
dc.description.versionostateczna wersja wydawcy
dc.description.volume76pl
dc.identifier.articleid882150pl
dc.identifier.eissn1899-0967pl
dc.identifier.issn1733-134Xpl
dc.identifier.projectROD UJ / OPpl
dc.identifier.urihttps://ruj.uj.edu.pl/xmlui/handle/item/57063
dc.identifier.weblinkhttp://archiwum.polradiol.com/abstract/index/idArt/882150pl
dc.languageengpl
dc.language.containerengpl
dc.rightsUdzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 4.0 Międzynarodowa*
dc.rights.licenceCC-BY-NC-ND
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/legalcode.pl*
dc.share.typeotwarte czasopismo
dc.subject.enindividual dosepl
dc.subject.endigital mammographypl
dc.subject.encomputed radiographypl
dc.subtypeOtherDocumentspl
dc.titleIndividual doses in screening mammography examinations performed with digital equipment in Polandpl
dc.title.journalPolish Journal of Radiologypl
dc.typeJournalArticlepl
dspace.entity.typePublication
dc.abstract.enpl
Background: The aim of this study was the estimation of individual doses for women participating in the mammography screening program in Poland, examined with digital systems. Material/Methods: Materials for this study included data obtained from 1800 exposures performed with the use of 9 different mammography units with digital systems. Exposure conditions (breast thickness after compression, high voltage, mAs value, the material of the anode and additional filter) were recorded during mammography screening examinations conducted for over one year in Poland. The collected data were used to determine individual doses in accordance with the method, which uses conversion factors for air kerma to the absorbed dose in the glandular tissue. The entrance air kerma for compressed breast exposure was determined on the basis of the measurements of the air kerma for the high-voltage value at which the exposure was performed, assuming a linear air kerma dependence on the tube load. Results: Individual doses established in 9 mammography facilities ranged from 0.40 mGy to 11.48 mGy for CR (computed radiography) systems and from 0.36 mGy to 5.80 mGy for DR (digital radiography) systems. The maximum value of the individual doses for CR systems was about 98% higher than for DR systems. The values of compressed breast thickness ranged from 2.0 cm to 7.6 cm for CR systems and from 2.1 cm to 11.4 cm for DR systems. The glandularity of breasts was from 9% to 100% for CR systems and from 3% to 97% for DR systems. The mean glandularity of breasts was 47% for CR systems and 29% for DR systems. Conclusions: One of the criteria for evaluating the suitability of digital systems in breast cancer screening in Poland should be the systematic evaluation of individual doses. In Poland, it is necessary to implement standardized methods of optimization of the exposure parameters for reducing individual doses to a minimum necessary to maintain the image quality.
dc.contributor.authorpl
Fabiszewska, Ewa
dc.contributor.authorpl
Grabska, Iwona
dc.contributor.authorpl
Pasicz, Katarzyna
dc.date.accessionpl
2018-09-18
dc.date.accessioned
2018-09-18T10:57:04Z
dc.date.available
2018-09-18T10:57:04Z
dc.date.issuedpl
2011
dc.date.openaccess
0
dc.description.accesstime
w momencie opublikowania
dc.description.additionalpl
Bibliogr. s. 79-80
dc.description.numberpl
4
dc.description.physicalpl
75-80
dc.description.version
ostateczna wersja wydawcy
dc.description.volumepl
76
dc.identifier.articleidpl
882150
dc.identifier.eissnpl
1899-0967
dc.identifier.issnpl
1733-134X
dc.identifier.projectpl
ROD UJ / OP
dc.identifier.uri
https://ruj.uj.edu.pl/xmlui/handle/item/57063
dc.identifier.weblinkpl
http://archiwum.polradiol.com/abstract/index/idArt/882150
dc.languagepl
eng
dc.language.containerpl
eng
dc.rights*
Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 4.0 Międzynarodowa
dc.rights.licence
CC-BY-NC-ND
dc.rights.uri*
http://creativecommons.org/licenses/by-nc-nd/4.0/legalcode.pl
dc.share.type
otwarte czasopismo
dc.subject.enpl
individual dose
dc.subject.enpl
digital mammography
dc.subject.enpl
computed radiography
dc.subtypepl
OtherDocuments
dc.titlepl
Individual doses in screening mammography examinations performed with digital equipment in Poland
dc.title.journalpl
Polish Journal of Radiology
dc.typepl
JournalArticle
dspace.entity.type
Publication
Affiliations

* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.

Views
1
Views per month
Downloads
fabiszewska_grabska_pasicz_individual_doses_in_screening_mammography_examinations_2011.pdf
23
fabiszewska_grabska_pasicz_individual_doses_in_screening_mammography_examinations_2011.odt
7