Retrospektywna analiza kryteriów dozymetrycznych w radioterapii dynamicznymi wiązkami łukowymi

master
dc.abstract.enThe purpose was to conduct the retrospective analysis of dosimetry criteria for treatment plan verification in intensity modulated arc therapy. Plans was realized between 09.11.2015 and 28.04.2017 in Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. The additional purpose was to judge an arbitrarily accepted dosimetry cryteria for the gamma parameter. This thesis include a detailed description of the radiotherapy treatment process, especially of the VMAT technique. The SPC (Statistical Process Control) method was used for the analysis. Plans have been divided on the grounds of the therapeutic area and also on account of the quarters. Analysis showed high quality and efficency for most areas. This confirms the validity of the criteria arbitrarily accepted in the Zakład Fizyki Medycznej, Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. The exception was the region of gynecological tumors, where the Cpml coefficient was 0.99. Additionally, only for head and neck location, the% GP average value has not exceeded 97%. In the third and fourth quarter of 2016 as well as in the first quarter of 2017, for all analyzed areas deterioration in quality and process performance was observed.pl
dc.abstract.plGłównym celem pracy było przeprowadzenie retrospektywnej analizy kryteriów dozymetrycznych w radioterapii dynamicznymi wiązkami łukowymi, dla planów realizowanych między 09.11.2015 a 28.04.2017 w Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. Jednym z zadań była również ocena zasadności arbitralnie przyjętych kryteriów dozymetrycznych, założonych dla współczynnika gamma. Praca zawiera szczegółowy opis procesu leczenia radioterapeutycznego, ze szczególnym uwzględnieniem techniki VMAT. Do przeprowadzenia analizy zastosowano metodę SPC (ang. Statistical Process Control). Dokonano podziału planów ze względu na obszar terapeutyczny a także ze względu na kwartały, w których były one realizowane. Analiza wykazała wysoką jakość oraz wydajność procesu dla większości obszarów terapeutycznych. Potwierdza to zasadności arbitralnie przyjętych w Zakładzie Fizyki Medycznej Instytutu im. Marii Skłodowskiej-Curie Oddział w Krakowie kryteriów założonych dla współczynnika gamma. Wyjątkiem był region nowotworów ginekologicznych, w którym współczynnik Cpml osiągnął wartość 0,99. Dodatkowo jedynie dla lokalizacji głowy i szyi wartość średnia %GP nie przekroczyła wartości 97%. W trzecim i czwartym kwartale 2016r. a także w pierwszym kwartale 2017r., dla wszystkich obszarów analizowanych zaobserwowano pogorszenie się jakości oraz wydajności procesu.pl
dc.affiliationWydział Fizyki, Astronomii i Informatyki Stosowanejpl
dc.areaobszar nauk ścisłychpl
dc.contributor.advisorStanek, Jan - 100112 pl
dc.contributor.advisorWaligórski, Michałpl
dc.contributor.authorŚwieczewski, Michałpl
dc.contributor.departmentbycodeUJK/WFAISpl
dc.contributor.reviewerStanek, Jan - 100112 pl
dc.contributor.reviewerStępień, Ewa - 161583 pl
dc.contributor.reviewerWaligórski, Michałpl
dc.date.accessioned2020-07-27T11:38:10Z
dc.date.available2020-07-27T11:38:10Z
dc.date.submitted2017-10-04pl
dc.fieldofstudyfizyka medycznapl
dc.identifier.apddiploma-119101-219994pl
dc.identifier.projectAPD / Opl
dc.identifier.urihttps://ruj.uj.edu.pl/xmlui/handle/item/224115
dc.languagepolpl
dc.subject.enquality assurance, SPC, volumetric modulated arc therapy, VMAT, radiotherapypl
dc.subject.plkontrola jakości, SPC, technika łukowa, VMAT, radioterapiapl
dc.titleRetrospektywna analiza kryteriów dozymetrycznych w radioterapii dynamicznymi wiązkami łukowymipl
dc.title.alternativeRetrospective analysis of dosimetry criteria for treatment plan verification in intensity modulated arc therapypl
dc.typemasterpl
dspace.entity.typePublication
dc.abstract.enpl
The purpose was to conduct the retrospective analysis of dosimetry criteria for treatment plan verification in intensity modulated arc therapy. Plans was realized between 09.11.2015 and 28.04.2017 in Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. The additional purpose was to judge an arbitrarily accepted dosimetry cryteria for the gamma parameter. This thesis include a detailed description of the radiotherapy treatment process, especially of the VMAT technique. The SPC (Statistical Process Control) method was used for the analysis. Plans have been divided on the grounds of the therapeutic area and also on account of the quarters. Analysis showed high quality and efficency for most areas. This confirms the validity of the criteria arbitrarily accepted in the Zakład Fizyki Medycznej, Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. The exception was the region of gynecological tumors, where the Cpml coefficient was 0.99. Additionally, only for head and neck location, the% GP average value has not exceeded 97%. In the third and fourth quarter of 2016 as well as in the first quarter of 2017, for all analyzed areas deterioration in quality and process performance was observed.
dc.abstract.plpl
Głównym celem pracy było przeprowadzenie retrospektywnej analizy kryteriów dozymetrycznych w radioterapii dynamicznymi wiązkami łukowymi, dla planów realizowanych między 09.11.2015 a 28.04.2017 w Centrum Onkologii Instytut im. Marii Skłodowskiej-Curie Oddział w Krakowie. Jednym z zadań była również ocena zasadności arbitralnie przyjętych kryteriów dozymetrycznych, założonych dla współczynnika gamma. Praca zawiera szczegółowy opis procesu leczenia radioterapeutycznego, ze szczególnym uwzględnieniem techniki VMAT. Do przeprowadzenia analizy zastosowano metodę SPC (ang. Statistical Process Control). Dokonano podziału planów ze względu na obszar terapeutyczny a także ze względu na kwartały, w których były one realizowane. Analiza wykazała wysoką jakość oraz wydajność procesu dla większości obszarów terapeutycznych. Potwierdza to zasadności arbitralnie przyjętych w Zakładzie Fizyki Medycznej Instytutu im. Marii Skłodowskiej-Curie Oddział w Krakowie kryteriów założonych dla współczynnika gamma. Wyjątkiem był region nowotworów ginekologicznych, w którym współczynnik Cpml osiągnął wartość 0,99. Dodatkowo jedynie dla lokalizacji głowy i szyi wartość średnia %GP nie przekroczyła wartości 97%. W trzecim i czwartym kwartale 2016r. a także w pierwszym kwartale 2017r., dla wszystkich obszarów analizowanych zaobserwowano pogorszenie się jakości oraz wydajności procesu.
dc.affiliationpl
Wydział Fizyki, Astronomii i Informatyki Stosowanej
dc.areapl
obszar nauk ścisłych
dc.contributor.advisorpl
Stanek, Jan - 100112
dc.contributor.advisorpl
Waligórski, Michał
dc.contributor.authorpl
Świeczewski, Michał
dc.contributor.departmentbycodepl
UJK/WFAIS
dc.contributor.reviewerpl
Stanek, Jan - 100112
dc.contributor.reviewerpl
Stępień, Ewa - 161583
dc.contributor.reviewerpl
Waligórski, Michał
dc.date.accessioned
2020-07-27T11:38:10Z
dc.date.available
2020-07-27T11:38:10Z
dc.date.submittedpl
2017-10-04
dc.fieldofstudypl
fizyka medyczna
dc.identifier.apdpl
diploma-119101-219994
dc.identifier.projectpl
APD / O
dc.identifier.uri
https://ruj.uj.edu.pl/xmlui/handle/item/224115
dc.languagepl
pol
dc.subject.enpl
quality assurance, SPC, volumetric modulated arc therapy, VMAT, radiotherapy
dc.subject.plpl
kontrola jakości, SPC, technika łukowa, VMAT, radioterapia
dc.titlepl
Retrospektywna analiza kryteriów dozymetrycznych w radioterapii dynamicznymi wiązkami łukowymi
dc.title.alternativepl
Retrospective analysis of dosimetry criteria for treatment plan verification in intensity modulated arc therapy
dc.typepl
master
dspace.entity.type
Publication
Affiliations

* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.

Views
17
Views per month
Views per city
Kielce
3
Dublin
2
Krakow
2
Wroclaw
2
Aleksandrów Łódzki
1
Brzozow
1
Chorowice
1
Lodz
1
Poznan
1
Szczecin
1

No access

No Thumbnail Available