Precise mass determination of single cell with cantilever-based microbiosensor system
pl
dc.type
JournalArticle
pl
dc.abstract.en
Having determined the mass of a single cell of brewer yeast Saccharomyces cerevisiae by means of a microcantilever-based biosensor Cantisens CSR-801 (Concentris, Basel, Switzerland), it was found that its dry mass is 47,65 $\pm$ 1,05 pg. Found to be crucial in this mass determination was the cell position along the length of the cantilever. Moreover, calculations including cells positions on the cantilever provide a threefold better degree of accuracy than those which assume uniform mass distribution. We have also examined the influence of storage time on the single cell mass. Our results show that after 6 months there is an increase in the average mass of a single yeast cell.
pl
dc.description.volume
12
pl
dc.description.number
11
pl
dc.identifier.doi
10.1371/journal.pone.0188388
pl
dc.identifier.eissn
1932-6203
pl
dc.title.journal
PLoS ONE
pl
dc.language.container
eng
pl
dc.affiliation
Wydział Fizyki, Astronomii i Informatyki Stosowanej : Instytut Fizyki im. Mariana Smoluchowskiego
pl
dc.subtype
Article
pl
dc.identifier.articleid
e0188388
pl
dc.rights.original
CC-BY; otwarte czasopismo; ostateczna wersja wydawcy; w momencie opublikowania; 0