Comparison between prospectively electrocardiogram-gated high-pitch mode and retrospectively electrocardiogram-gated mode for dual-source CT coronary angiography

2015
journal article
article
dc.abstract.enBACKGROUND: We aimed to report the image quality, relationship between heart rate and image quality, amount of contrast agent given to the patients and radiation doses in coronary CT angiography (CTA) obtained by using high-pitch prospectively ECG-gated "Flash Spiral" technique (method A) or retrospectively ECG-gated technique (method B) using 128×2-slice dual-source CT. MATERIAL AND METHODS: A total of 110 patients who were evaluated with method A and method B technique with a 128×2-detector dual-source CT device were included in the study. Patients were divided into three groups based on their heart rates during the procedure, and a relationship between heart rate and image quality were evaluated. The relationship between heart rate, gender and radiation dose received by the patients was compared. RESULTS: A total of 1760 segments were evaluated in terms of image quality. Comparison of the relationship between heart rate and image quality revealed a significant difference between heart rate <60 beats/min group and >75 beats/min group whereas <60 beats/min and 60-75 beats/min groups did not differ significantly. The average effective dose for coronary CTA was calculated as 1.11 mSv (0.47-2.01 mSv) for method A and 8.22 mSv (2.19-12.88 mSv) for method B. CONCLUSIONS: Method A provided high quality images with doses as low as <1 mSv in selected patients who have low heart rates with a high negative predictive value to rule out coronary artery disease. Although method B increases the amount of effective dose, it provides high diagnostic quality images for patients who have a high heart rate and arrhythmia which makes it is difficult to obtain images.pl
dc.contributor.authorKoplay, Mustafapl
dc.contributor.authorCelik, Mahmutpl
dc.contributor.authorAvci, Ahmetpl
dc.contributor.authorErdogan, Hasanpl
dc.contributor.authorDemir, Kenanpl
dc.contributor.authorSivri, Mesutpl
dc.contributor.authorNayman, Alaaddinpl
dc.date.accessioned2017-07-18T12:29:40Z
dc.date.available2017-07-18T12:29:40Z
dc.date.issued2015pl
dc.date.openaccess0
dc.description.accesstimew momencie opublikowania
dc.description.additionalBibliogr. s. 567-568pl
dc.description.physical561-568pl
dc.description.versionostateczna wersja wydawcy
dc.description.volume80pl
dc.identifier.doi10.12659/PJR.895232pl
dc.identifier.eissn1899-0967pl
dc.identifier.issn1733-134Xpl
dc.identifier.urihttp://ruj.uj.edu.pl/xmlui/handle/item/42933
dc.languageengpl
dc.language.containerengpl
dc.rightsUdzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska*
dc.rights.licenceCC-BY-NC-ND
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode*
dc.share.typeotwarte czasopismo
dc.subject.encoronary angiographypl
dc.subject.enheart ratepl
dc.subject.enimage enhancementpl
dc.subject.enmultidetector computed tomographypl
dc.subject.enradiation dosagepl
dc.subject.enradiologypl
dc.subtypeArticlepl
dc.titleComparison between prospectively electrocardiogram-gated high-pitch mode and retrospectively electrocardiogram-gated mode for dual-source CT coronary angiographypl
dc.title.journalPolish Journal of Radiologypl