Oversampling in the computed tomography measurements applied for bone structure studies as a method of spatial resolution improvement
computed tomography
image processing
image resolution
oversampling
Bibliogr. s. 18
Background: Our purpose was to check the potential ability of oversampling as a method for computed tomography axial resolution improvement. The method of achieving isotropic and fine resolution, when the scanning system is characterized by anisotropic resolutions is proposed. In case of typical clinical system the axial resolution is much lower than the planar one. The idea relies on the scanning with a wide overlapping layers and subsequent resolution recovery on the level of scanning step. Material/Methods: Simulated three-dimensional images, as well as the real microtomographic images of rat femoral bone were used in proposed solution tests. Original high resolution images were virtually scanned with a wide beam and a small step in order to simulate the real measurements. The low resolution image series were subsequently processed in order to back to the original fine one. Original, virtually scanned and recovered images resolutions were compared with the use of modulation transfer function (MTF). Results/Conclusions: A good ability of oversampling as a method for the resolution recovery was showed. It was confirmed by comparing the resolving powers after and before resolution recovery. The MTF analysis showed resolution improvement. The resolution improvement was achieved but the image noise raised considerably, which is clearly visible on image histograms. Despite this disadvantage the proposed method can be successfully used in practice, especially in the trabecular bone studies because of high contrast between trabeculae and intertrabecular spaces.
dc.abstract.en | Background: Our purpose was to check the potential ability of oversampling as a method for computed tomography axial resolution improvement. The method of achieving isotropic and fine resolution, when the scanning system is characterized by anisotropic resolutions is proposed. In case of typical clinical system the axial resolution is much lower than the planar one. The idea relies on the scanning with a wide overlapping layers and subsequent resolution recovery on the level of scanning step. Material/Methods: Simulated three-dimensional images, as well as the real microtomographic images of rat femoral bone were used in proposed solution tests. Original high resolution images were virtually scanned with a wide beam and a small step in order to simulate the real measurements. The low resolution image series were subsequently processed in order to back to the original fine one. Original, virtually scanned and recovered images resolutions were compared with the use of modulation transfer function (MTF). Results/Conclusions: A good ability of oversampling as a method for the resolution recovery was showed. It was confirmed by comparing the resolving powers after and before resolution recovery. The MTF analysis showed resolution improvement. The resolution improvement was achieved but the image noise raised considerably, which is clearly visible on image histograms. Despite this disadvantage the proposed method can be successfully used in practice, especially in the trabecular bone studies because of high contrast between trabeculae and intertrabecular spaces. | pl |
dc.affiliation | Wydział Lekarski : Zakład Biofizyki | pl |
dc.affiliation | Wydział Fizyki, Astronomii i Informatyki Stosowanej : Instytut Fizyki im. Mariana Smoluchowskiego | pl |
dc.contributor.author | Tatoń, Grzegorz - 133654 | pl |
dc.contributor.author | Rokita, Eugeniusz - 133291 | pl |
dc.contributor.author | Rok, Tomasz | pl |
dc.contributor.author | Beckmann, Felix | pl |
dc.date.accessioned | 2016-05-25T13:49:10Z | |
dc.date.available | 2016-05-25T13:49:10Z | |
dc.date.issued | 2012 | pl |
dc.date.openaccess | 0 | |
dc.description.accesstime | w momencie opublikowania | |
dc.description.additional | Bibliogr. s. 18 | pl |
dc.description.number | 2 | pl |
dc.description.physical | 14-18 | pl |
dc.description.version | ostateczna wersja wydawcy | |
dc.description.volume | 77 | pl |
dc.identifier.doi | 10.12659/PJR.882965 | pl |
dc.identifier.eissn | 1899-0967 | pl |
dc.identifier.issn | 1733-134X | pl |
dc.identifier.uri | http://ruj.uj.edu.pl/xmlui/handle/item/27041 | |
dc.language | eng | pl |
dc.language.container | eng | pl |
dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska | * |
dc.rights.licence | CC-BY-NC-ND | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode | * |
dc.share.type | otwarte czasopismo | |
dc.subject.en | computed tomography | pl |
dc.subject.en | image processing | pl |
dc.subject.en | image resolution | pl |
dc.subject.en | oversampling | pl |
dc.subtype | Article | pl |
dc.title | Oversampling in the computed tomography measurements applied for bone structure studies as a method of spatial resolution improvement | pl |
dc.title.journal | Polish Journal of Radiology | pl |
dc.type | JournalArticle | pl |
dspace.entity.type | Publication |
* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.
Open Access