Background: Recanalization occurs occasionally, following coil embolization of pulmonary arteriovenous malformations (PAVM), and can lead to ischemic stroke; therefore re-embolization is important. A 1.9-Fr. no-taper microcatheter that can be inserted into a 2.7-Fr. microcatheter (named the triaxial system) has recently become available, and contributes to super-selective catheterization for small or tortuous vessels. The aim of this study was to evaluate the usefulness of re-embolization for recanalization of PAVM using the triaxial system. Material and Methods: Recanalization was diagnosed in 8 patients with 13 PAVMs between June 2011 and November 2012, and re-embolization was attempted with a conventional microcatheter at first in all 13 PAVMs. However, in three of them it failed with the conventional microcatheter, and then the system was exchanged to the triaxial system. Thus, re-embolization using the triaxial system was performed in 3 PAVMs of 3 female patients, with a median age of 63 years (range, 46–73 years). We assessed technical success, complications, and outcome. Results: The disappearance of recanalization was confirmed by angiography in all re-embolization procedures (technical success rate was 100%). Re-embolization was then successfully achieved inside the original coils, and no branch artery of normal lung tissue was embolized. There were no complications related with this procedure. The blood flow of recanalization was decreased in all cases in a follow-up of 27-33 months (median, 31). Conclusions: Triaxial system appears to be useful for recanalization of PAVM, especially in difficult cases with a conventional system.
cris.lastimport.wos | 2024-04-09T20:52:19Z | |
dc.abstract.en | Background: Recanalization occurs occasionally, following coil embolization of pulmonary arteriovenous malformations (PAVM), and can lead to ischemic stroke; therefore re-embolization is important. A 1.9-Fr. no-taper microcatheter that can be inserted into a 2.7-Fr. microcatheter (named the triaxial system) has recently become available, and contributes to super-selective catheterization for small or tortuous vessels. The aim of this study was to evaluate the usefulness of re-embolization for recanalization of PAVM using the triaxial system. Material and Methods: Recanalization was diagnosed in 8 patients with 13 PAVMs between June 2011 and November 2012, and re-embolization was attempted with a conventional microcatheter at first in all 13 PAVMs. However, in three of them it failed with the conventional microcatheter, and then the system was exchanged to the triaxial system. Thus, re-embolization using the triaxial system was performed in 3 PAVMs of 3 female patients, with a median age of 63 years (range, 46–73 years). We assessed technical success, complications, and outcome. Results: The disappearance of recanalization was confirmed by angiography in all re-embolization procedures (technical success rate was 100%). Re-embolization was then successfully achieved inside the original coils, and no branch artery of normal lung tissue was embolized. There were no complications related with this procedure. The blood flow of recanalization was decreased in all cases in a follow-up of 27-33 months (median, 31). Conclusions: Triaxial system appears to be useful for recanalization of PAVM, especially in difficult cases with a conventional system. | pl |
dc.contributor.author | Shimohira, Masashi | pl |
dc.contributor.author | Hashizume, Takuya | pl |
dc.contributor.author | Kawai, Tatsuya | pl |
dc.contributor.author | Muto, Masahiro | pl |
dc.contributor.author | Ohta, Kengo | pl |
dc.contributor.author | Suzuki, Kazushi | pl |
dc.contributor.author | Shibamoto, Yuta | pl |
dc.date.accessioned | 2017-07-06T11:40:25Z | |
dc.date.available | 2017-07-06T11:40:25Z | |
dc.date.issued | 2015 | pl |
dc.date.openaccess | 0 | |
dc.description.accesstime | w momencie opublikowania | |
dc.description.additional | Bibliogr. s. 70-71 | pl |
dc.description.physical | 67-71 | pl |
dc.description.version | ostateczna wersja wydawcy | |
dc.description.volume | 80 | pl |
dc.identifier.doi | 10.12659/PJR.892378 | pl |
dc.identifier.eissn | 1899-0967 | pl |
dc.identifier.issn | 1733-134X | pl |
dc.identifier.uri | http://ruj.uj.edu.pl/xmlui/handle/item/42448 | |
dc.language | eng | pl |
dc.language.container | eng | pl |
dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska | * |
dc.rights.licence | CC-BY-NC-ND | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode | * |
dc.share.type | otwarte czasopismo | |
dc.subject.en | catheterization | pl |
dc.subject.en | embolization | pl |
dc.subject.en | therapeutic | pl |
dc.subject.en | pulmonary artery | pl |
dc.subtype | Article | pl |
dc.title | Triaxial system in re-embolization for recanalization of pulmonary arteriovenous malformations | pl |
dc.title.journal | Polish Journal of Radiology | pl |
dc.type | JournalArticle | pl |
dspace.entity.type | Publication |
* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.
Views
4
Views per month
Views per city
Downloads
Open Access