Evaluation of doses and image quality in mammography with screen-film, CR, and DR detectors : application of the ACR phantom
mammography
quality control
radiation dosage
Bibliogr. s. 390-391
BACKGROUND: Different methods of image quality evaluation are routinely used for analogue and digital mammography systems in Poland. In the present study, image quality for several screen-film (SF), computed radiography (CR), and fully digital (DR) mammography systems was compared directly with the use of the ACR mammography accreditation phantom. MATERIAL AND METHODS: Image quality and mean glandular doses were measured and compared for 47 mammography systems in the Mazovia Voivodeship in Poland, including 26 SF systems, 12 CR systems, and 9 DR systems. The mean glandular dose for the breast simulated by 4.5 cm of PMMA was calculated with methods described in the "European guidelines for quality assurance in breast cancer screening and diagnosis". Visibility of the structures in the image (fibers, microcalcifications, and masses) was evaluated with the mammographic accreditation ACR phantom. RESULTS: Image quality for DR systems was significantly higher than for SF and CR systems. Several SF systems failed to pass the image quality tests because of artifacts. The doses were within acceptable limits for all of the systems, but the doses for the CR systems were significantly higher than for the SF and DR systems. CONCLUSIONS: The best image quality, at a reasonably low dose, was observed for the DR systems. The CR systems are capable of obtaining the same image quality as the SF systems, but only at a significantly higher dose. The ACR phantom can be routinely used to evaluate image quality for all types of mammographic systems.
dc.abstract.en | BACKGROUND: Different methods of image quality evaluation are routinely used for analogue and digital mammography systems in Poland. In the present study, image quality for several screen-film (SF), computed radiography (CR), and fully digital (DR) mammography systems was compared directly with the use of the ACR mammography accreditation phantom. MATERIAL AND METHODS: Image quality and mean glandular doses were measured and compared for 47 mammography systems in the Mazovia Voivodeship in Poland, including 26 SF systems, 12 CR systems, and 9 DR systems. The mean glandular dose for the breast simulated by 4.5 cm of PMMA was calculated with methods described in the "European guidelines for quality assurance in breast cancer screening and diagnosis". Visibility of the structures in the image (fibers, microcalcifications, and masses) was evaluated with the mammographic accreditation ACR phantom. RESULTS: Image quality for DR systems was significantly higher than for SF and CR systems. Several SF systems failed to pass the image quality tests because of artifacts. The doses were within acceptable limits for all of the systems, but the doses for the CR systems were significantly higher than for the SF and DR systems. CONCLUSIONS: The best image quality, at a reasonably low dose, was observed for the DR systems. The CR systems are capable of obtaining the same image quality as the SF systems, but only at a significantly higher dose. The ACR phantom can be routinely used to evaluate image quality for all types of mammographic systems. | pl |
dc.contributor.author | Ślusarczyk-Kacprzyk, Wioletta | pl |
dc.contributor.author | Skrzyński, Witold | pl |
dc.contributor.author | Fabiszewska, Ewa | pl |
dc.date.accessioned | 2017-05-09T09:50:49Z | |
dc.date.available | 2017-05-09T09:50:49Z | |
dc.date.issued | 2016 | pl |
dc.date.openaccess | 0 | |
dc.description.accesstime | w momencie opublikowania | |
dc.description.additional | Bibliogr. s. 390-391 | pl |
dc.description.physical | 386-391 | pl |
dc.description.version | ostateczna wersja wydawcy | |
dc.description.volume | 81 | pl |
dc.identifier.doi | 10.12659/PJR.897304 | pl |
dc.identifier.eissn | 1899-0967 | pl |
dc.identifier.issn | 1733-134X | pl |
dc.identifier.uri | http://ruj.uj.edu.pl/xmlui/handle/item/40235 | |
dc.language | eng | pl |
dc.language.container | eng | pl |
dc.rights | Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne - Bez utworów zależnych 3.0 Polska | * |
dc.rights.licence | CC-BY-NC-ND | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/pl/legalcode | * |
dc.share.type | otwarte czasopismo | |
dc.subject.en | mammography | pl |
dc.subject.en | quality control | pl |
dc.subject.en | radiation dosage | pl |
dc.subtype | Article | pl |
dc.title | Evaluation of doses and image quality in mammography with screen-film, CR, and DR detectors : application of the ACR phantom | pl |
dc.title.journal | Polish Journal of Radiology | pl |
dc.type | JournalArticle | pl |
dspace.entity.type | Publication |
* The migration of download and view statistics prior to the date of April 8, 2024 is in progress.
Views
1
Views per month
Views per city
Downloads
Open Access